In mass vaccination, there are various reasons for high pressure in the system.

For example, the operator can hit the cartilage, there might be particles in the vaccine that block the vaccine flow, the user might be using a single injection head instead of a dual head, and so on.

In order to lower the pressure, After high-pressure detection, the system will repump the max dose (0.6ml).

In the High-Presusre procedure, the system will help direct you to the problem by asking a few questions.

The first two times the system detects High-Pressure, it will give the operator a warning alert and allow continue working, in case the reason is hitting the cartilage or a small blockage that fixed itself.

Control unit notification

Control unit LED

Hand unit LEDs

Duration

Comments

High-Pressure
Detection
Continue working

Yellow

Yellow light indication on the right side.
Red light
indication on the left side.

Red light- one second
The alert will disappear after 20 seconds or if the user presses ok

The injection was not completed

The third time, the system detects high pressure, it will try to release the pressure by returning the injector to the home position.

  • If it does succeed in returning to the home position and there is still pressure, this notification will appear:

Control unit notification

Control unit LED

Hand unit LEDs

Duration

Comments

High-Pressure Partial blockage Check needle, pump Press OK

Red

Red light indication on the right side

The message stays on the screen until the user clicks OK.

The user won’t be able to inject in the meantime.

  • If it doesn't succeed- and we are using:

  1. A Dual system

  2. The high pressure is in the secondary unit or in both primary and secondary.

    The following message will appear:

Control unit notification

Control unit LED

Hand unit LEDs

Duration

Comments

High-Pressure

Verify the use of Dual

Injection Head

Then press ok

Red

Red light indication on the right side

The message stays on the screen until the user clicks OK.

The user won’t be able to inject in the meantime.

If one of the above conditions is false or if they were true and the user replaced the injection head, the system still detects high pressure.

The following message will appear:

Control unit notification

Control unit LED

Hand unit LEDs

Duration

Comments

High-Pressure

Replace the needle

Then press ok

Red

Red light indication on the right side

The message stays on the screen until the user clicks OK.

The user won’t be able to inject in the meantime.

The needle counter will be reset automatically when the user presses OK.

After the user presses OK, the system will try to release pressure by returning the injector to the home position.

If it doesn't succeed, The following message will appear:

Control unit notification

Control unit LED

Hand unit LEDs

Duration

Comments

High-Pressure Error

Shutdown the system

Red

Red light indication on the right side

The message stays on the screen until the user shuts down the system.

The user won’t be able to inject in the meantime.