Our newest firmware release for the pHi-Tech devices includes several new features:
✨When the bottle is almost empty
The hand unit will turn on an orange blinking light if 20% or fewer injections are left.
It will stop once you reset and replace the bottle.Please Click here for more information on how to set the almost bottle empty alert.
The temperature sampling has been adjusted
Instead of sampling once per injection, the system will sample ten times at the end of each injection. It will then display the average of these ten samples.
The early needle pull mechanism is improved
Now, it will be detected if the needle is inside for less than the injection time with a more accurate calculation.✨Numbering the Alerts for sorting and getting easier indications for the operators-
1 | Replace Injector |
2 | Replace Needle |
3 | Replace Injection Head |
4 | Bottle Empty |
5 | Cleaning & Disinfection Needed |
6 | Bottle almost empty |
7 | Early Neddle pull |
8 | BLE error |
9 | Hand Unit communication error, Shutdown the system |
10 | Primary / Secondary comm error Shutdown the system |
11 | Low Battery |
12 | Critical battery level |
13 | Cleaning & Disinfection temp low |
14 | Cleaning & Disinfection temp high |
15 | Low vaccine temp |
16 | High vaccine temp |
21 | Actuator failure check the needle press OK |
22 | High-Pressure Error- Shutdown the system |
23 | Cleaning & Disinfection Extreme temp high |
24 | Clean injector |
27 | Motor stuck shutdown the system |
28 | Optical Sensor error |
29 | High-Pressure Verify the use of dual injection head |
31 | High-Pressure- Replace the needle |
32 | High Pressure Detection- Continue working |
33 | Injection Counter reached max display continue working |
34 | High-Pressure Partial blockage Check needle, pump Press OK |
5. Bug fixes
6. ✨A New Bootloader Version 1.3
📢 Please Note A Bootloader Version can't be updated by FOTA. * To get this bootloader version- please contact your pHi-Tech regional specialist.
|
The Firmware is not automatically deleted when the device is in FOTA mode
You need to Approve the process in the Mobile App,
and Only then, the firmware will be deleted and the upgrade will start.
* Once the process starts, you have to finish the upgrade to be able to use the device.
If there is no firmware when turning the device on, it will automatically go to FOTA mode. There is no need to press the down button while turning on the device.
7. The new firmware version is now available in French as well.
To upgrade to the X.4.14 firmware version- you need to perform the FOTA process.
If you are having any issues please contact pHi-Tech support.