Firmware version X.4.14

X.4.14 with bootloader version 1.3

Hadas Movshovitz avatar
Written by Hadas Movshovitz
Updated over 4 months ago

Our newest firmware release for the pHi-Tech devices includes several new features:

  1. ✨When the bottle is almost empty
    The hand unit will turn on an orange blinking light if 20% or fewer injections are left.


    It will stop once you reset and replace the bottle.

    Please Click here for more information on how to set the almost bottle empty alert.

  2. The temperature sampling has been adjusted
    Instead of sampling once per injection, the system will sample ten times at the end of each injection. It will then display the average of these ten samples.

  3. The early needle pull mechanism is improved
    Now, it will be detected if the needle is inside for less than the injection time with a more accurate calculation.

  4. Numbering the Alerts for sorting and getting easier indications for the operators-

1

Replace Injector

2

Replace Needle

3

Replace Injection Head

4

Bottle Empty

5

Cleaning & Disinfection Needed

6

Bottle almost empty

7

Early Neddle pull

8

BLE error

9

Hand Unit communication error, Shutdown the system

10

Primary / Secondary comm error Shutdown the system

11

Low Battery

12

Critical battery level

13

Cleaning & Disinfection temp low

14

Cleaning & Disinfection temp high

15

Low vaccine temp

16

High vaccine temp

21

Actuator failure check the needle press OK

22

High-Pressure Error- Shutdown the system

23

Cleaning & Disinfection Extreme temp high

24

Clean injector

27

Motor stuck shutdown the system

28

Optical Sensor error

29

High-Pressure Verify the use of dual injection head

31

High-Pressure- Replace the needle

32

High Pressure Detection- Continue working

33

Injection Counter reached max display continue working

34

High-Pressure Partial blockage Check needle, pump Press OK

5. Bug fixes

6. A New Bootloader Version 1.3

📢 Please Note

A Bootloader Version can't be updated by FOTA.

This Version is only available for New systems (received from July 2024)
Or systems upgraded by a pHi-Tech representative.

* To get this bootloader version- please contact your pHi-Tech regional specialist.

  • The Firmware is not automatically deleted when the device is in FOTA mode
    You need to Approve the process in the Mobile App,
    and Only then, the firmware will be deleted and the upgrade will start.

    * Once the process starts, you have to finish the upgrade to be able to use the device. ​

  • If there is no firmware when turning the device on, it will automatically go to FOTA mode. There is no need to press the down button while turning on the device.

7. The new firmware version is now available in French as well.


To upgrade to the X.4.14 firmware version- you need to perform the FOTA process.

If you are having any issues please contact pHi-Tech support.