Ir al contenido principal
Calibración

test

Osmany Moncada avatar
Escrito por Osmany Moncada
Actualizado hace más de una semana

  1. Switch on the Control Unit


2. Navigate to the Calibration screen.

3. Click on the OK button to Enable the Calibration process.

4. Take a syringe in your hands and point the needle on an angle inside.
With an automatic Injection Head- Press the trigger on the Hand Unit and pull
back the Injection Head.
With a manual Injection Head, just press the trigger.
The system will automatically deliver ten doses.

5. Check the injected material volume in the syringe.
Correct Volume
If the dosage in the syringe is the correct amount, press the down arrow to
reach the Save field and click OK and continue to step 6.

Incorrect Volume
If the dosage in the syringe is not correct, go to the 'Inj.Vol' field and click OK.

When you see the full arrow, use the up and down arrows to set the volume
that is currently in the syringe.

When done, Click OK.

Go to the Save field and click OK.

6. Press OK to disable the Calibration mode.

  • When using a Dual system, after calibrating the first vaccine bottle, repeat the process for the second vaccine bottle.

¿Ha quedado contestada tu pregunta?